FDA Enforcement Class II Terminated

Wallace Sure-Pro¿ Two Stage Embryo Replacement Catheter for difficult transfer with Stylet PPS623 Sure-Pro¿ Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra¿ Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the one-stage procedure, the clinician pushes the inner catheter through the outer sheath and loads the embryo prior to performing the embryo transfer into the uterus. In the two-stage procedure, the clinician places the outer sheath through the cervical canal into the uterus, then loads the embryo in the inner catheter, and advances the inner catheter through the outer sheath to perform the embryo transfer into the uterus.

Recall: Z-1243-2013 · Reported May 15, 2013

Enforcement

Recall Number
Z-1243-2013
Event ID
64705
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Smiths Medical ASD, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 15, 2013
Initiation Date
March 28, 2013
Classification Date
May 3, 2013
Termination Date
May 19, 2021
Address
10 Bowman Dr, N/A, Keene, NH, 03431-5043, United States

Description

Wallace Sure-Pro¿ Two Stage Embryo Replacement Catheter for difficult transfer with Stylet PPS623 Sure-Pro¿ Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra¿ Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the one-stage procedure, the clinician pushes the inner catheter through the outer sheath and loads the embryo prior to performing the embryo transfer into the uterus. In the two-stage procedure, the clinician places the outer sheath through the cervical canal into the uterus, then loads the embryo in the inner catheter, and advances the inner catheter through the outer sheath to perform the embryo transfer into the uterus.

Reason

Unable to pass catheter through outer sheath during embryo transfer

Code Info

Lot Code: 2000103, 2025752, 2161556, 2165854, 2170929, 2304190, 2363603

Distribution

Worldwide distribution: USA (Nationwide) and countries of: Arab Emirates (AE) Ireland (IE), Albania, Israel (IL), Argentina (AR),Italy (IT) Armenia (AM) , Japan (JP), Austria (AT) Jordan (JO), Bahrain (BH) Kuwait (KW) Belgium (BE), Lebanon (LB), Belarus , Netherlands (NL), Brazil (BR),Nigeria (NG), Bulgaria (BG) , Pakistan (PK), Canada (CA) Poland (PL), China (CN) Romania (RO), Columbia (CO), Russia (RU), Cyprus(CY) , Saudi Arabia (SA) Czech (CZ) , Singapore (SG), Denmark (OK), South Korea (KR), Egypt (EG), Spain (ES), Finland (FI), Sri Lanka (LK),Germany (DE) Turkey (TR) Ghana (GH) , Uganda (UG), Great Britain (GB) Ukraine (UA), Greece (GR) Uruguay (UY), and Hong Kong (HK).

Quantity

4910 units