Elekta Precise Treatment Table

Recall: Z-1238-2015 · Reported March 25, 2015

Enforcement

Recall Number
Z-1238-2015
Event ID
70458
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 25, 2015
Initiation Date
January 29, 2015
Classification Date
March 19, 2015
Termination Date
November 5, 2021
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227, United States

Description

Elekta Precise Treatment Table

Reason

It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.

Code Info

Precise Treatment Table Serial Numbers: (124001-125803, 125807-880, 125883-899, 125902-917, 125920-949, 125952-963, 125966-1256453, 126158-170, 126174-208, 126212-230, 126232-261, 126269-281, 126293-309,123318-330, 126338-126373, 126375-397, 126400-133999, 213000+)

Distribution

Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland and Syria.

Quantity

529 US 2951 Worldwide