FDA Enforcement Class II Ongoing

DELFIA¿ hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.

Recall: Z-1237-2019 · Reported May 8, 2019

Enforcement

Recall Number
Z-1237-2019
Event ID
82064
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PerkinElmer Life and Analytical Sciences, Wallac, OY
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2019
Initiation Date
January 10, 2019
Classification Date
April 26, 2019
Address
Mustionkatu 6, P.O. Box 10, Turku, N/A, N/A, Finland

Description

DELFIA¿ hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.

Reason

hCG analyte concentrations may be decreased up to 15- 20% at the lower measuring range of hCG concentrations.

Code Info

Kit Lot Numbers: 664891, 666471, 666604 Pack Lot Numbers: 1066489101, 1066647101,1066660401 UDI: (01)06438147009176(17)190430(10)664891 (01)06438147009176(17)190430(10)666471 (01)06438147009176(17)190430(10)666604

Distribution

Worldwide distribution - US Nationwide distribution. and countries of Austria, Bolivia, UK, Switzerland,Thailand, Finland.

Quantity

31 kits (10 in US, 21 in ROW)