FDA Enforcement Class II Ongoing

CellMek SPS Sample Preparation System, REF: C44603, with software

Recall: Z-1235-2026 · Reported February 11, 2026

Enforcement

Recall Number
Z-1235-2026
Event ID
98207
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 11, 2026
Initiation Date
December 23, 2025
Classification Date
February 2, 2026
Address
11800 Sw 147th Ave, N/A, Miami, FL, 33196-2500, United States

Description

CellMek SPS Sample Preparation System, REF: C44603, with software

Reason

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Code Info

UDI-DI: 15099590750312. Software Version 2.3.90.0

Distribution

Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.

Quantity

50