CellMek SPS Sample Preparation System, REF: C44603, with software
Recall: Z-1235-2026
·
Reported February 11, 2026
Enforcement
- Recall Number
- Z-1235-2026
- Event ID
- 98207
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Beckman Coulter, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 11, 2026
- Initiation Date
- December 23, 2025
- Classification Date
- February 2, 2026
- Address
- 11800 Sw 147th Ave, N/A, Miami, FL, 33196-2500, United States
Description
CellMek SPS Sample Preparation System, REF: C44603, with software
Reason
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Code Info
UDI-DI: 15099590750312. Software Version 2.3.90.0
Distribution
Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.
Quantity
50