FDA Enforcement Class II Terminated

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

Recall: Z-1235-2023 · Reported March 15, 2023

Enforcement

Recall Number
Z-1235-2023
Event ID
91795
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Sensus Healthcare, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 15, 2023
Initiation Date
February 10, 2023
Classification Date
March 8, 2023
Termination Date
January 31, 2024
Address
851 Broken Sound Pkwy Nw Ste 215, Boca Raton, FL, 33487-3635, United States

Description

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

Reason

When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.

Code Info

Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272

Distribution

US Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.

Quantity

23 units