FDA Enforcement Class II Terminated

Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461

Recall: Z-1235-2019 · Reported May 8, 2019

Enforcement

Recall Number
Z-1235-2019
Event ID
82521
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2019
Initiation Date
December 7, 2018
Classification Date
April 26, 2019
Termination Date
April 27, 2020
Address
333 Coney St, N/A, East Walpole, MA, 02032-1516, United States

Description

Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461

Reason

Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).

Code Info

Lot Number/Exp. Date: kit lots ending in 317 and lower 72259303 12/13/2018 87903305 1/12/2019 94908307 2/11/2019 96439307 2/11/2019 16824309 3/14/2019 25822309 3/14/2019 28942311 4/27/2019 55978313 5/27/2019 70389317 6/29/2019

Distribution

Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana

Quantity

1489 kits