FDA Enforcement Class II Terminated

ACE Trochanteric Nail System STER L H TROC NAIL 320X9X125 STER L H TROC NAIL 340X9X125 STER L H TROC NAIL 360X9X125 STER L H TROC NAIL 380X9X125 STER L H TROC NAIL 400X9X125 STER L H TROC NAIL 420X9X125 STER L H TROC NAIL 440X9X125 STER L H TROC NAIL 460X9X125 STER L H TROC NAIL 320X11X125 STER L H TROC NAIL 340X11X125 STER L H TROC NAIL 360X11X125 STER L H TROC NAIL 380X11X125 STER L H TROC NAIL 400X11X125 STER L H TROC NAIL 420X11X125 STER L H TROC NAIL 440X11X125 STER L H TROC NAIL 460X11X125 STER L H TROC NAIL 320X13X125 STER L H TROC NAIL 340X13X125 STER L H TROC NAIL 360X13X125 STER L H TROC NAIL 380X13X125 STER L H TROC NAIL 400X13X125 STER L H TROC NAIL 420X13X125 STER L H TROC NAIL 440X13X125 STER L H TROC NAIL 460X13X125 STER R H TROC NAIL 320X9X125 STER R H TROC NAIL 340X9X125 STER R H TROC NAIL 360X9X125 STER R H TROC NAIL 380X9X125 STER R H TROC NAIL 400X9X125 STER R H TROC NAIL 420X9X125 STER R H TROC NAIL 440X9X125 STER R H TROC NAIL 460X9X125 STER R H TROC NAIL 320X11X125 STER R H TROC NAIL 340X11X125 STER R H TROC NAIL 360X11X125 STER R H TROC NAIL 380X11X125 STER R H TROC NAIL 400X11X125 STER R H TROC NAIL 420X11X125 STER R H TROC NAIL 440X11X125 STER R H TROC NAIL 460X11X125 STER R H TROC NAIL 320X13X125 STER R H TROC NAIL 340X13X125 STER R H TROC NAIL 360X13X125 STER R H TROC NAIL 380X13X125 STER R H TROC NAIL 400X13X125 STER R H TROC NAIL 420X13X125 STER R H TROC NAIL 440X13X125 STER R H TROC NAIL 460X13X125 STER L H TROC NAIL 320X11X130 STER L H TROC NAIL 320X9X130 STER L H TROC NAIL 340X9X130 STER L H TROC NAIL 360X9X130 STER L H TROC NAIL 380X9X130 STER L H TROC NAIL 400X9X130 STER L H TROC NAIL 420X9X130 STER L H TROC NAIL 440X9X130 STER L H TROC NAIL 460X9X130 STER L H TROC NAIL 340X11X130 STER L H TROC NAIL 360X11X130 STER L H TROC NAIL 380X11X130 STER L H TROC NAIL 400X11X130 STER L H TROC NAIL 420X11X130 STER L H TROC NAIL 440X11X130 STER L H TROC NAIL 460X11X130 STER L H TROC NAIL 320X13X130 STER L H TROC NAIL 340X13X130 STER L H TROC NAIL 360X13X130 STER L H TROC NAIL 380X13X130 STER L H TROC NAIL 400X13X130 STER L H TROC NAIL 420X13X130 STER L H TROC NAIL 440X13X130 STER L H TROC NAIL 460X13X130 STER R H TROC NAIL 320X9X130 STER R H TROC NAIL 340X9X130 STER R H TROC NAIL 360X9X130 STER R H TROC NAIL 380X9X130 STER R H TROC NAIL 400X9X130 STER R H TROC NAIL 420X9X130 STER R H TROC NAIL 440X9X130 STER R H TROC NAIL 460X9X130 STER R H TROC NAIL 320X11X130 STER R H TROC NAIL 340X11X130 STER R H TROC NAIL 360X11X130 STER R H TROC NAIL 380X11X130 STER R H TROC NAIL 400X11X130 STER R H TROC NAIL 420X11X130 STER R H TROC NAIL 440X11X130 STER R H TROC NAIL 320X13X130 STER R H TROC NAIL 340X13X130 STER R H TROC NAIL 360X13X130 STER R H TROC NAIL 380X13X130 STER R H TROC NAIL 400X13X130 STER R H TROC NAIL 420X13X130 STER R H TROC NAIL 440X13X130 STER R H TROC NAIL 460X13X130 STER L H TROC NAIL 320X9X135 STER L H TROC NAIL 340X9X135 STER L H TROC NAIL 360X9X135 STER L H TROC NAIL 380X9X135 STER L H TROC NAIL 400X9X135 STER L H TROC NAIL 420X9X135 STER L H TROC NAIL 440X9X135 STER L H TROC NAIL 460X9X135 STER L H TROC NAIL 320X11X135 STER L H TROC NAIL 340X11X135 STER L H TROC NAIL 360X11X135 STER L H TROC NAIL 380X11X135 STER L H TROC NAIL 400X11X135 STER L H TROC NAIL 420X11X135 STER L H TROC NAIL 440X11X135 STER L H TROC NAIL 460X11X135 STER L H TROC NAIL 320X13X135 STER L H TROC NAIL 340X13X135 STER L H TROC NAIL 360X13X135 STER L H TROC NAIL 380X13X135 STER L H TROC NAIL 400X13X135 STER L H TROC NAIL 420X13X135 STER L H TROC NAIL 440X13X135 STER L H TROC NAIL 460X13X135 STER R H TROC NAIL 320X13X135 STER R H TROC NAIL 340X13X135 STER R H TROC NAIL 360X13X135 STER R H TROC NAIL 380X13X135 STER R H TROC NAIL 400X13X135 STER R H TROC NAIL 420X13X135 STER R H TROC NAIL 440X13X135 STER R H TROC NAIL 460X13X135 STER R H TROC NAIL 320X9X135 STER R H TROC NAIL 340X9X135 STER R H TROC

Recall: Z-1235-2018 · Reported April 4, 2018

Enforcement

Recall Number
Z-1235-2018
Event ID
79408
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 4, 2018
Initiation Date
December 20, 2017
Classification Date
March 27, 2018
Termination Date
April 22, 2020
Address
56 E Bell Dr, Warsaw, IN, 46582-6989, United States

Description

ACE Trochanteric Nail System STER L H TROC NAIL 320X9X125 STER L H TROC NAIL 340X9X125 STER L H TROC NAIL 360X9X125 STER L H TROC NAIL 380X9X125 STER L H TROC NAIL 400X9X125 STER L H TROC NAIL 420X9X125 STER L H TROC NAIL 440X9X125 STER L H TROC NAIL 460X9X125 STER L H TROC NAIL 320X11X125 STER L H TROC NAIL 340X11X125 STER L H TROC NAIL 360X11X125 STER L H TROC NAIL 380X11X125 STER L H TROC NAIL 400X11X125 STER L H TROC NAIL 420X11X125 STER L H TROC NAIL 440X11X125 STER L H TROC NAIL 460X11X125 STER L H TROC NAIL 320X13X125 STER L H TROC NAIL 340X13X125 STER L H TROC NAIL 360X13X125 STER L H TROC NAIL 380X13X125 STER L H TROC NAIL 400X13X125 STER L H TROC NAIL 420X13X125 STER L H TROC NAIL 440X13X125 STER L H TROC NAIL 460X13X125 STER R H TROC NAIL 320X9X125 STER R H TROC NAIL 340X9X125 STER R H TROC NAIL 360X9X125 STER R H TROC NAIL 380X9X125 STER R H TROC NAIL 400X9X125 STER R H TROC NAIL 420X9X125 STER R H TROC NAIL 440X9X125 STER R H TROC NAIL 460X9X125 STER R H TROC NAIL 320X11X125 STER R H TROC NAIL 340X11X125 STER R H TROC NAIL 360X11X125 STER R H TROC NAIL 380X11X125 STER R H TROC NAIL 400X11X125 STER R H TROC NAIL 420X11X125 STER R H TROC NAIL 440X11X125 STER R H TROC NAIL 460X11X125 STER R H TROC NAIL 320X13X125 STER R H TROC NAIL 340X13X125 STER R H TROC NAIL 360X13X125 STER R H TROC NAIL 380X13X125 STER R H TROC NAIL 400X13X125 STER R H TROC NAIL 420X13X125 STER R H TROC NAIL 440X13X125 STER R H TROC NAIL 460X13X125 STER L H TROC NAIL 320X11X130 STER L H TROC NAIL 320X9X130 STER L H TROC NAIL 340X9X130 STER L H TROC NAIL 360X9X130 STER L H TROC NAIL 380X9X130 STER L H TROC NAIL 400X9X130 STER L H TROC NAIL 420X9X130 STER L H TROC NAIL 440X9X130 STER L H TROC NAIL 460X9X130 STER L H TROC NAIL 340X11X130 STER L H TROC NAIL 360X11X130 STER L H TROC NAIL 380X11X130 STER L H TROC NAIL 400X11X130 STER L H TROC NAIL 420X11X130 STER L H TROC NAIL 440X11X130 STER L H TROC NAIL 460X11X130 STER L H TROC NAIL 320X13X130 STER L H TROC NAIL 340X13X130 STER L H TROC NAIL 360X13X130 STER L H TROC NAIL 380X13X130 STER L H TROC NAIL 400X13X130 STER L H TROC NAIL 420X13X130 STER L H TROC NAIL 440X13X130 STER L H TROC NAIL 460X13X130 STER R H TROC NAIL 320X9X130 STER R H TROC NAIL 340X9X130 STER R H TROC NAIL 360X9X130 STER R H TROC NAIL 380X9X130 STER R H TROC NAIL 400X9X130 STER R H TROC NAIL 420X9X130 STER R H TROC NAIL 440X9X130 STER R H TROC NAIL 460X9X130 STER R H TROC NAIL 320X11X130 STER R H TROC NAIL 340X11X130 STER R H TROC NAIL 360X11X130 STER R H TROC NAIL 380X11X130 STER R H TROC NAIL 400X11X130 STER R H TROC NAIL 420X11X130 STER R H TROC NAIL 440X11X130 STER R H TROC NAIL 320X13X130 STER R H TROC NAIL 340X13X130 STER R H TROC NAIL 360X13X130 STER R H TROC NAIL 380X13X130 STER R H TROC NAIL 400X13X130 STER R H TROC NAIL 420X13X130 STER R H TROC NAIL 440X13X130 STER R H TROC NAIL 460X13X130 STER L H TROC NAIL 320X9X135 STER L H TROC NAIL 340X9X135 STER L H TROC NAIL 360X9X135 STER L H TROC NAIL 380X9X135 STER L H TROC NAIL 400X9X135 STER L H TROC NAIL 420X9X135 STER L H TROC NAIL 440X9X135 STER L H TROC NAIL 460X9X135 STER L H TROC NAIL 320X11X135 STER L H TROC NAIL 340X11X135 STER L H TROC NAIL 360X11X135 STER L H TROC NAIL 380X11X135 STER L H TROC NAIL 400X11X135 STER L H TROC NAIL 420X11X135 STER L H TROC NAIL 440X11X135 STER L H TROC NAIL 460X11X135 STER L H TROC NAIL 320X13X135 STER L H TROC NAIL 340X13X135 STER L H TROC NAIL 360X13X135 STER L H TROC NAIL 380X13X135 STER L H TROC NAIL 400X13X135 STER L H TROC NAIL 420X13X135 STER L H TROC NAIL 440X13X135 STER L H TROC NAIL 460X13X135 STER R H TROC NAIL 320X13X135 STER R H TROC NAIL 340X13X135 STER R H TROC NAIL 360X13X135 STER R H TROC NAIL 380X13X135 STER R H TROC NAIL 400X13X135 STER R H TROC NAIL 420X13X135 STER R H TROC NAIL 440X13X135 STER R H TROC NAIL 460X13X135 STER R H TROC NAIL 320X9X135 STER R H TROC NAIL 340X9X135 STER R H TROC

Reason

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

Code Info

915209320 915209340 915209360 915209380 915209400 915209420 915209440 915209460 915211320 915211340 915211360 915211380 915211400 915211420 915211440 915211460 915213320 915213340 915213360 915213380 915213400 915213420 915213440 915213460 916209320 916209340 916209360 916209380 916209400 916209420 916209440 916209460 916211320 916211340 916211360 916211380 916211400 916211420 916211440 916211460 916213320 916213340 916213360 916213380 916213400 916213420 916213440 916213460 925200320 925209320 925209340 925209360 925209380 925209400 925209420 925209440 925209460 925211340 925211360 925211380 925211400 925211420 925211440 925211460 925213320 925213340 925213360 925213380 925213400 925213420 925213440 925213460 926209320 926209340 926209360 926209380 926209400 926209420 926209440 926209460 926211320 926211340 926211360 926211380 926211400 926211420 926211440 926213320 926213340 926213360 926213380 926213400 926213420 926213440 926213460 935209320 935209340 935209360 935209380 935209400 935209420 935209440 935209460 935211320 935211340 935211360 935211380 935211400 935211420 935211440 935211460 935213320 935213340 935213360 935213380 935213400 935213420 935213440 935213460 936213320 936213340 936213360 936213380 936213400 936213420 936213440 936213460 936309320 936309340 936309360 936309380 936309400 936309420 936309440 936309460 936311320 936311340 936311360 936311380 936311400 936311420 936311440 936311460

Distribution

USA (nationwide)

Quantity

3454