Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke
Recall: Z-1234-2025
·
Reported March 5, 2025
Enforcement
- Recall Number
- Z-1234-2025
- Event ID
- 96158
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Scientia Vascular, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 5, 2025
- Initiation Date
- January 2, 2025
- Classification Date
- February 26, 2025
- Address
- 2460 S 3270 W, N/A, West Valley City, UT, 84119-1116, United States
Description
Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke
Reason
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
Code Info
Catalog Number: SC038-127-001 UDI-DI code: 00818075011025 Lot Numbers: 030471 030317 030345 030376 030427 030451
Distribution
U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.