SiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 10492399
Enforcement
- Recall Number
- Z-1234-2019
- Event ID
- 82521
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 8, 2019
- Initiation Date
- December 7, 2018
- Classification Date
- April 26, 2019
- Termination Date
- April 27, 2020
- Address
- 333 Coney St, N/A, East Walpole, MA, 02032-1516, United States
Description
SiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 10492399
Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).
Lot Number/Exp. Date: (Kit Lots ending in 316 or lower) 68750302 12/13/2018 68751302 12/13/2018 83578304 1/12/2019 91382304 1/12/2019 4642306 2/11/2019 8116306 2/11/2019 19098308 3/14/2019 22635308 3/14/2019 33640310 4/27/2019 39133310 4/27/2019 55625312 5/27/2019 55626312 5/27/2019 56285312 5/27/2019 69206316 6/29/2019 73001316 6/29/2019
Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana
10414 kits