FDA Enforcement Class II Ongoing

Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Recall: Z-1233-2023 · Reported March 15, 2023

Enforcement

Recall Number
Z-1233-2023
Event ID
91773
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Elekta Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 15, 2023
Initiation Date
February 28, 2023
Classification Date
March 7, 2023
Address
1450 Beale St Ste 205, Saint Charles, MO, 63303, United States

Description

Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Reason

Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

Code Info

Software Build: 5.11.00, 5.11.01, 5.11.02, 5.11.03; UDI Numbers: (01)00858164002169(10)5.11.00, (01)00858164002169(10)5.11.01, (01)00858164002169(10)5.11.02, (01)00858164002169(10)5.11.03

Distribution

Nationwide distribution in the US. International distribution to Albania, Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Occupied, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Zimbabwe.

Quantity

2,020 units