FDA Enforcement
Class II
Terminated
Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10700444
Recall: Z-1233-2020
·
Reported February 26, 2020
Enforcement
- Recall Number
- Z-1233-2020
- Event ID
- 84612
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 26, 2020
- Initiation Date
- December 18, 2019
- Classification Date
- February 14, 2020
- Termination Date
- June 25, 2020
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005, United States
Description
Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10700444
Reason
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
Code Info
UDI Number: 00842768036262 Lots: All lots
Distribution
US Nationwide
Quantity
11,725 units