FDA Enforcement Class II Terminated

Stryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Recall: Z-1233-2013 · Reported May 15, 2013

Enforcement

Recall Number
Z-1233-2013
Event ID
64682
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 15, 2013
Initiation Date
March 27, 2013
Classification Date
May 3, 2013
Termination Date
September 24, 2013
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

Stryker Irrigator Reusable Tips: 5mm X 45cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Reason

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Code Info

Manufacturer Part Number: 250-070-488

Distribution

Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.

Quantity

147,921 - all sizes