FDA Enforcement
Class II
Ongoing
Cardiosave Rescue. Model Number: 0998-00-0800-83.
Recall: Z-1232-2023
·
Reported March 15, 2023
Enforcement
- Recall Number
- Z-1232-2023
- Event ID
- 91642
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Datascope Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 15, 2023
- Initiation Date
- February 7, 2023
- Classification Date
- March 6, 2023
- Address
- 1300 Macarthur Blvd, Mahwah, NJ, 07430-2052, United States
Description
Cardiosave Rescue. Model Number: 0998-00-0800-83.
Reason
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
Code Info
Model Number: 0998-00-0800-83. UDI-DI: 10607567108407. All serial numbers.
Distribution
Nationwide US distribution. Worldwide international distribution.
Quantity
47 units (44 US, 3 OUS)