Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, and 0998-00-0800-55
Enforcement
- Recall Number
- Z-1229-2023
- Event ID
- 91645
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Datascope Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 15, 2023
- Initiation Date
- February 7, 2023
- Classification Date
- March 6, 2023
- Address
- 1300 Macarthur Blvd, Mahwah, NJ, 07430-2052, United States
Description
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, and 0998-00-0800-55
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
All Serial Numbers. Model Number: 0998-00-0800-31, UDI-DI:10607567109053; Model Number: 0998-00-0800-32, UDI-DI: 10607567111117; Model Number: 0998-00-0800-33, UDI-DI: 10607567109008; Model Number: 0998-00-0800-34, UDI-DI: 10607567111940; Model Number: 0998-00-0800-35, UDI-DI: 10607567109107; Model Number: 0998-00-0800-45, UDI-DI: 10607567108421; Model Number: 0998-00-0800-52, UDI-DI: 10607567108438; Model Number: 0998-00-0800-53, UDI-DI: 10607567108391; Model Number: 0998-00-0800-55, UDI-DI: 10607567108414; Model Number: 0998-00-0800-65, UDI-DI: 10607567113432
US Nationwide. Global Distribution.
8909 units