FDA Enforcement Class II Ongoing

CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 51mm Material Number: 66-351 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).

Recall: Z-1223-2021 · Reported March 17, 2021

Enforcement

Recall Number
Z-1223-2021
Event ID
87251
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 17, 2021
Initiation Date
January 15, 2021
Classification Date
March 11, 2021
Address
375 River Park Cir, Marquette, MI, 49855-1781, United States

Description

CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 51mm Material Number: 66-351 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).

Reason

Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision

Code Info

Batch Number: 333021 333022 338840 341489 343445 346886 351527 351528 351528 351528 351528 351528 351528 358459 358459 358459 358459 358459 358459 362405 363444 364766 364766 369197 369197 370268 377838 377838 379930 379930 379930 379930 379931 379931 385407

Distribution

US Nationwide distribution.