FDA Enforcement Class II Terminated

Vital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S, Product usage: The devices are intended for use in patient therapy requiring long-term vascular access for infusion therapy and/or blood sampling.

Recall: Z-1223-2018 · Reported April 4, 2018

Enforcement

Recall Number
Z-1223-2018
Event ID
79425
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Vascular Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 4, 2018
Initiation Date
November 28, 2017
Classification Date
March 27, 2018
Termination Date
October 3, 2018
Address
1186 Montgomery Ln, Vandergrift, PA, 15690-6065, United States

Description

Vital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S, Product usage: The devices are intended for use in patient therapy requiring long-term vascular access for infusion therapy and/or blood sampling.

Reason

Non-coring needle provided with the Cook Vital-Port¿ Vascular Assess System (Vital-Port) may cut or dislodge a core or sliver of material from the Vital-Port septum when the non-coring needle is inserted into the Vital-Port. This needle is used on the initial implant of the Vital-Port. Vital-Port products that have been successfully placed in patients are not impacted by this recall.

Code Info

IPI, P-S

Distribution

The devices have been distributed within the United States and the following countries: United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Bulgaria, Bermuda, Brazil, Canada, Switzerland, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Spain, Finland, Faroe Islands, France, United Kingdom, Guatemala, Hong Kong, Hungary, Indonesia, Ireland, Iceland, Italy, Jamaica, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Latvia, Morocco, Macedonia, Mexico, Netherlands, New Zealand, Oman, Panama, Poland, Portugal, Serbia, Saudi Arabia, Sweden, San Marino, El Salvador, Turkey, Trinidad & Tobago, Ukraine, Uruguay, Venezuela, Virgin Islands, Kosovo, China, and South Africa.

Quantity

31,006 Units