Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.

Recall: Z-1217-2014 · Reported March 26, 2014

Enforcement

Recall Number
Z-1217-2014
Event ID
67486
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Exactech, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 26, 2014
Initiation Date
January 6, 2014
Classification Date
March 18, 2014
Termination Date
August 4, 2014
Address
2320 NW 66th Ct, Gainesville, FL, 32653-1630, United States

Description

Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.

Reason

The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.

Code Info

size 3.5mm x 10mm, Catalog #05-000-20-3510, Lot Number #35917001. size 3.5mm x 14mm, Catalog #05-000-20-3514, Lot Number 35917002 and 40023002. size 3.5mm x 20mm, Catalog #05-000-20-3520, Lot Number 35917003. size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917004. size 4.0mm x 14mm, Catalog #05-000-20-4014, Lot Number 35917005. size 4.0mm x 20mm, Catalog #05-000-20-4020, Lot Number 35917006. size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917007. 1/4 Shank, size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917008. 1/4 Shank, size 3.5mm x 30mm, Catalog #05-000-20-3530, Lot Number 35917009. 1/4 Shank, size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917010. 1/4 Shank, size 4.0mm x 30mm, Catalog #05-000-20-4030, Lot Number 35917011.

Distribution

USA Distribution including the states of FL, NY, OR, IL, PA, MO and NJ.

Quantity

242