FDA Enforcement
Class II
Ongoing
LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
Recall: Z-1216-2023
·
Reported March 15, 2023
Enforcement
- Recall Number
- Z-1216-2023
- Event ID
- 91646
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 15, 2023
- Initiation Date
- February 2, 2023
- Classification Date
- March 3, 2023
- Address
- 4100 Hamline Ave N, Saint Paul, MN, 55112-5700, United States
Description
LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
Reason
Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..
Code Info
UDI-DI: 00802526613876
Distribution
US Nationwide distribution to the healthcare facilities and healthcare providers are located in CA, CO, DE, FL, GA, ID, IL, IN, IA, KY, ME, MD, MA, NV, NH, NJ, NY, NC, OH, PA, RI, SC, TN TX, VA, WA, WV, and WI.
Quantity
1,531 patient data records