FDA Enforcement
Class III
Terminated
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Recall: Z-1216-2017
·
Reported February 22, 2017
Enforcement
- Recall Number
- Z-1216-2017
- Event ID
- 76293
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 22, 2017
- Initiation Date
- May 5, 2016
- Classification Date
- February 16, 2017
- Termination Date
- February 2, 2018
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Reason
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Code Info
Lot Number: 4MD048, 5DD026, 5ED071, 5GD004, 5JD058, 5LD061
Distribution
Distributed throughout the United States
Quantity
13,147 units