FDA Enforcement Class II Ongoing

CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 32mm Material Number: 66-232 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).

Recall: Z-1214-2021 · Reported March 17, 2021

Enforcement

Recall Number
Z-1214-2021
Event ID
87251
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 17, 2021
Initiation Date
January 15, 2021
Classification Date
March 11, 2021
Address
375 River Park Cir, Marquette, MI, 49855-1781, United States

Description

CervAlign Anterior Cervical Plate System-Cervalign, 2Level, 32mm Material Number: 66-232 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).

Reason

Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision

Code Info

Batch Number: 330973 330975 330976 338830 338831 344099 344100 350516 350517 356798 358456 361803 361804 364749 369336 370393 377188 379018 379542 379543 379544 380874

Distribution

US Nationwide distribution.