FDA Enforcement Class II Terminated

Product Brand Name(s): Alere Triage¿ TOX+MTD Drug Screen, Catalog Number 94400, Lots W56013B and W56035B. 25 individually pouched devices in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen is a fluorescence immunoassay to be used with the Alere Triage¿ Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses. This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/ Mass Spectrometry/ Mass Spectrometry (LC/MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods. A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.

Recall: Z-1214-2014 · Reported March 26, 2014

Enforcement

Recall Number
Z-1214-2014
Event ID
67529
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Alere San Diego, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 26, 2014
Initiation Date
February 5, 2014
Classification Date
March 17, 2014
Termination Date
October 27, 2014
Address
9975 Summers Ridge Rd, San Diego, CA, 92121-2997, United States

Description

Product Brand Name(s): Alere Triage¿ TOX+MTD Drug Screen, Catalog Number 94400, Lots W56013B and W56035B. 25 individually pouched devices in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen is a fluorescence immunoassay to be used with the Alere Triage¿ Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses. This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/ Mass Spectrometry/ Mass Spectrometry (LC/MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods. A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.

Reason

Alere San Diego, Inc. is initiating a recall of the Alere Triage¿ TOX Drug Screen, Catalog Number 94400, Lots W56013B and W56035B because Alere's investigation has shown that these lots have a frequency of false negative results for PCP (Phencyclidine) that is higher than the frequency shown in the package insert. If you have questions about previously reported results.

Code Info

Catalog Number 94400, Lots W56013B and W56035B

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Taiwan.

Quantity

305 kits (239 kits of W56013B, 66 kits of W56035B)