Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

Recall: Z-1213-2014 · Reported March 26, 2014

Enforcement

Recall Number
Z-1213-2014
Event ID
67517
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mar Cor Purification
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
March 26, 2014
Initiation Date
June 26, 2013
Classification Date
March 14, 2014
Termination Date
January 22, 2015
Address
14550 28th Ave N, Minneapolis, MN, 55447-4817, United States

Description

Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

Reason

High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.

Code Info

Serial Numbers: 1294627, 1298383, 1298386, 1299352, 1299353, 1299354, 1299355, 1299356, 1299486, 1299487, 1299488, 1299489, 1299503, 1299572, 1299712, 1299713, 1299716, 1299718, 1299719, 1299720, 1299740, 1299742, 1299777, 1299779, 1299780, 1299831, 1299835, 1299850, 1299867, 1299868, 1299895, 1299897, 1299992, 1299993, 1299994, 1299996, 1299997, 1300138, 1300224, 1300248, 1300363, 1300364, 1300367, 1300368, 1300558, 1300560, 1300561, 1300562, 1300564, 1300565, 1300621, 1300649, 1300650, 1300700, 1300701, 1301081, 1301116, 1301360, 1301361, 1301386, 1301888, 1301990, 1301992, 1301997, 1302030, 1302032, 1302149, 1302214, 1302226, 1302342, 1302345, 1302346, 1302385, 1302430, 1302447, 1302448, 1302452, 1302456, 1302464, 1302517, 1302555, 1302558, 1302642, 1302696, 1302706, 1302708, 1302748, 1302767, 1302768, 1302810, 1302813, 1302922, 1302923, 1302924, 1302925, 1302926, 1302927, 1302987, 1303005, 1303108, 1303315, 1303316, 1303346, 1303348, 1303351, 1303352, 1303505, 1303516, 1303532, 1303533, 1303534, 1303557, 1303558, 1303625, 1303626, 1303689, 1303690, 1303693, 1303702, 1303711, 1303768, 1303781, 1303799, 1303806, 1303821, 1303872, 1303877, 1303908, 1303920, 1303975, 1303976, 1304070, 1304076, 1304077, 1304081, 1304163, 1304166, 1304167, 1304170, 1304173, 1304174, 1304194, 1304196, 1304203, 1304273, 1304293, 1304294, 1304297, 1304333, 1304340, 1304343, 1304344, 1304346, 1304349, 1304376, 1304381, 1304501, 1304502, 1304503, 1304507, 1304508, 1304551, 1304585, 1304587, 1304724, 1304725, 1304775, 1304778, 1304779, 1304829, 1304870, 1304902, 1304903, 1304929, 1305006, 1305007, 1305038, 1305040, 1305041, 1305087, 1305088, 1305089, 1305131, 1305133, 1305210, 1305238, 1305239, 1305355, 1305363, 1305608, 1305609, 1305650, 1305651, 1305655, 1305698, 1305700, 1305701, 1305717, 1305732, 1305735, 1305750, 1305751, 1305752, 1305795, 1305928, 1305931, 1305932, 1305975, 1305995, 1306007, 1306042, 1306046, 1306078, 1306125, 1306154, 1306178, 1306209, 1306215, 1306226, 1306259, 1306322, 1306341, 1306428, 1306429, 1306430, 1306431, 1306432, 1306654, 1306662, 1306663, 1306664, 1306665.

Distribution

USA Nationwide Distribution including Puerto Rico.

Quantity

232 units