FDA Enforcement Class II Terminated

COMPASS SW Version 3.1, Catalog Number CS10-100, medical linear accelerator, radiological

Recall: Z-1212-2015 · Reported March 4, 2015

Enforcement

Recall Number
Z-1212-2015
Event ID
70545
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Iba Dosimetry Gmbh
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 4, 2015
Initiation Date
February 2, 2015
Classification Date
February 26, 2015
Termination Date
August 18, 2015
Address
Bahnhofstr. 5, N/A, Schwarzenbruck, N/A, N/A, Germany

Description

COMPASS SW Version 3.1, Catalog Number CS10-100, medical linear accelerator, radiological

Reason

Error in the software. During internal tests of the current development version of the Compass SW it was found that dose reconstruction for DMLC plans when the jaws move or mlc leaves reverse during beam on will not be scaled correctly for all control points.

Code Info

Model : SW Version 3.1 Lot or Serial Number : n.a

Distribution

Worldwide Distribution - US Distribution to the states of AZ and OK., and the countries of Germany, Austria, France, UK, Poland, Finland, The Netherlands, Spain, Turkey, Russia, and South Africa, Algeria, North and South America, Mexico, Brazil, Venezuela, Colombia, Panama, Asia, China, Japan, Philippines, Korea, India, Singapore, Thailand, Taiwan, Hong Kong, Malaysia and Guam.

Quantity

124 units