FDA Enforcement
Class II
Ongoing
Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.
Recall: Z-1210-2023
·
Reported March 15, 2023
Enforcement
- Recall Number
- Z-1210-2023
- Event ID
- 91504
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- DiaSorin Molecular LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 15, 2023
- Initiation Date
- January 30, 2023
- Classification Date
- March 3, 2023
- Address
- 11331 Valley View St, Cypress, CA, 90630-5366, United States
Description
Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.
Reason
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Code Info
UDI-DI: 20816101026962. Product REF/Lot/Reaction Mix REF/Lot: MOL2655/13648N/MOL2656/13649N, MOL2655/13882N/MOL2656/15164N, MOL2655/15279N/MOL2656/15280N
Distribution
US Nationwide distribution including in the states of NJ, ME, ND, IL, MI, FL, GA, AZ, MA, CA, MO, SD, MN, PA, OH, VT, NE, WI, CO, DC, VA, TX, NH, IN, IA, MD, NC, AL, NY, KY, WV, RI, KS, NM, OK, WA, SC, CT, TN, NV, LA.
Quantity
4142