FDA Enforcement Class II Terminated

Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1. AT-1 sets and kits (9005104 AT-1 Autotransfusion set, 9108504 ATF 120 Fast Start Kit, and 9108494 ATF 40 Fast Start Kits); 2. AT-3 Autotransfusion sets and kits (Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits AT-3) Product Usage: The Fresenius Kabi CATSmart device (Continuous auto transfusion System) is indicated for the processing of autologous shed blood collected interaoperatively and postoperatively to obtain washed packed red blood cells for reinfusion. The CATSmart Auto transfusion System is designed to reduce the need for allogeneic blood transfusion by re-infusing a patient s own lost blood during surgery. The set is used as a disposable component of the autotransfusion system during surgery and/or treatment of trauma to remove waste and improve the safety of autologous re-infused blood (auto transfusion). This is a single-use device.

Recall: Z-1209-2019 · Reported May 1, 2019

Enforcement

Recall Number
Z-1209-2019
Event ID
82395
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fenwal Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 1, 2019
Initiation Date
March 5, 2019
Classification Date
April 23, 2019
Termination Date
October 5, 2020
Address
3 Corporate Dr Ste 300, N/A, Lake Zurich, IL, 60047-8930, United States

Description

Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1. AT-1 sets and kits (9005104 AT-1 Autotransfusion set, 9108504 ATF 120 Fast Start Kit, and 9108494 ATF 40 Fast Start Kits); 2. AT-3 Autotransfusion sets and kits (Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits AT-3) Product Usage: The Fresenius Kabi CATSmart device (Continuous auto transfusion System) is indicated for the processing of autologous shed blood collected interaoperatively and postoperatively to obtain washed packed red blood cells for reinfusion. The CATSmart Auto transfusion System is designed to reduce the need for allogeneic blood transfusion by re-infusing a patient s own lost blood during surgery. The set is used as a disposable component of the autotransfusion system during surgery and/or treatment of trauma to remove waste and improve the safety of autologous re-infused blood (auto transfusion). This is a single-use device.

Reason

Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the centrifuge tubing line making contact with the washing chamber. The contact may result in excessive wear of the centrifuge tubing or in extreme cases a puncture of the tubing.

Code Info

Product Code: 9005901; Lot Numbers beginning with GM or H. Serial Numbers: 6CTA0209, 6CTA0210, 6CTA0211, 6CTA0212, 6CTA0213, 6CTA0214, 6CTA0215, 6CTA0216, 6CTA0217, 6CTA0218, 6CTA0219, 6CTA0221, 6CTA0222, 6CTA0223, 6CTA0224, 6CTA0225, 6CTA0226, 6CTA0227, 6CTA0228, 6CTA0260, 6CTA0261, 6CTA0262, 6CTA0263, 6CTA0264, 6CTA0265, 6CTA0266, 6CTA0267, 6CTA0268, 6CTA0269, 6CTA0300, 6CTA0301, 6CTA0302, 6CTA0303, 6CTA0304, 6CTA0305, 6CTA0306, 6CTA0307, 6CTA0308, 6CTA0309, 7CTA0328, 7CTA0329, 7CTA0330, 7CTA0331, 7CTA0332, 7CTA0334, 7CTA0335, 7CTA0336, 7CTA0337, 7CTA0358, 7CTA0359, 7CTA0360, 7CTA0361, 7CTA0362, 7CTA0363, 7CTA0364, 7CTA0365, 7CTA0366, 7CTA0367, 7CTA0368, 7CTA0369, 7CTA0370, 7CTA0371, 7CTA0372, 7CTA0373, 7CTA0374, 7CTA0375, 7CTA0376, 7CTA0377, 7CTA0378, 7CTA0379, 7CTA0380, 7CTA0381, 7CTA0392, 7CTA0393, 7CTA0394, 7CTA0395, 7CTA0396, 7CTA0397, 7CTA0398, 7CTA0399, 7CTA0401, 7CTA0402, 7CTA0403, 7CTA0404, 7CTA0405, 7CTA0406, 7CTA0409, 7CTA0410, 7CTA0411, 7CTA0412, 7CTA0413, 7CTA0414, 7CTA0433, 7CTA0434, 7CTA0435, 7CTA0436, 7CTA0437, 7CTA0438, 7CTA0439, 7CTA0440, 7CTA0472, 7CTA0473, 7CTA0474, 7CTA0475, 7CTA0476, 7CTA0477, 7CTA0478, 7CTA0479, 7CTA0480, 7CTA0481, 7CTA0482, 7CTA0483, 7CTA0484, 7CTA0485, 7CTA0486, 7CTA0487, 7CTA0488, 7CTA0489, 7CTA0509, 7CTA0510, 7CTA0511, 7CTA0512, 7CTA0513, 7CTA0514, 7CTA0515, 7CTA0516, 7CTA0517, 7CTA0518, 7CTA0519, 7CTA0520, 7CTA0521, 7CTA0522, 7CTA0523, 7CTA0548, 7CTA0549, 7CTA0550, 7CTA0551, 7CTA0552, 7CTA0553, 7CTA0554, 7CTA0556, 7CTA0557, 7CTA0558, 7CTA0580, 7CTA0581, 7CTA0582, 7CTA0583, 7CTA0584, 7CTA0585, 7CTA0586, 7CTA0587, 7CTA0588, 7CTA0589, 7CTA0590, 7CTA0591, 7CTA0647, 7CTA0648, 7CTA0649, 7CTA0650, 7CTA0651, 7CTA0652, 7CTA0653, 7CTA0654, 7CTA0655, 7CTA0656, 7CTA0657, 7CTA0658, 7CTA0659, 7CTA0660, 7CTA0661, 7CTA0662, 7CTA0663, 7CTA0664, 7CTA0665, 7CTA0666, 7CTA0687, 7CTA0688, 7CTA0689, 7CTA0690, 7CTA0691, 7CTA0692, 7CTA0696, 7CTA0697, 7CTA0698, 7CTA0699, 7CTA0700, 7CTA0701, 7CTA0702, 7CTA0715, 7CTA0716, 7CTA0717, 7CTA0719, 7CTA0720, 7CTA0721, 7CTA0722, 7CTA0723, 7CTA0724, 7CTA0735, 7CTA0736, 7CTA0737, 7CTA0738, 7CTA0739, 7CTA0740, 8CTA0741, 8CTA0742, 8CTA0748, 8CTA0749, 8CTA0754, 8CTA0911, 8CTA0912, 8CTA0913, 8CTA0914, 8CTA0915, 8CTA0916, 8CTA0917, 8CTA0918, 8CTA0919, 8CTA0920, 8CTA0921, 8CTA0922, 8CTA0923, 8CTA0924, 8CTA0925, 8CTA0926, 8CTA0927, 8CTA0928, 8CTA0929, 8CTA0930, 8CTA0931, 8CTA0932, 8CTA0933, 8CTA0934, 8CTA0935, 8CTA0936, 8CTA0937, 8CTA0938, 8CTA0939, 8CTA0940, 8CTA0942, 8CTA0943, 8CTA0944, 8CTA0945, 8CTA0946, 8CTA0947, 8CTA0948, 8CTA0949, 8CTA0950, 8CTA0951, 8CTA0952, 8CTA0953, 8CTA0954, 8CTA0955, 8CTA0956, 8CTA0957, 8CTA0958, 8CTA0959, 8CTA0960, 8CTA0961, 8CTA0962, 8CTA0963, 8CTA0964, 8CTA0965, 8CTA0966, 8CTA0967, 8CTA0968, 8CTA0969, 8CTA0970, 8CTA0971, 8CTA0972, 8CTA0973, 8CTA0974, 8CTA0975, 8CTA0976, 8CTA0977, 8CTA0978, 8CTA0979, 8CTA0980, 8CTA0981, 8CTA0982, 8CTA0983, 8CTA0984, 8CTA0985, 8CTA0986, 8CTA0987, 8CTA0988, 8CTA0989, 8CTA0990, 8CTA0991, 8CTA0992, 8CTA0993, 8CTA0994, 8CTA0996, 8CTA0997, 8CTA0998, 8CTA0999, 8CTA1000, 8CTA1001, 8CTA1002, 8CTA1003, 8CTA1004, 8CTA1005, 8CTA1006, 8CTA1007, 8CTA1008, 8CTA1009, 8CTA1010, 8CTA1011, 8CTA1032, 8CTA1033, 8CTA1037, 8CTA1039, 8CTA1040, 8CTA1041, 8CTA1042, 8CTA1053, 8CTA1057, 8CTA1056, 8CTA1055, 8CTA1054, 8CTA1051, 6CTA0220, 8CTA1050, 8CTA1049, 8CTA1048, 8CTA1047, 8CTA1046, 8CTA1045, 7CTA0333, 8CTA1044, 8CTA1043, 8CTA0995, 8CTA1024, 8CTA1025, 8CTA1026, 8CTA1027, 8CTA1028, 8CTA1029, 8CTA1030, 8CTA1031, 8CTA1034, 8CTA1035, 8CTA1036, 8CTA1038

Distribution

Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and PR, and country of Canada.

Quantity

342 devices