FDA Enforcement
Class II
Ongoing
IntelePACS (Image Fusion Module) - InteleViewer
Recall: Z-1208-2025
·
Reported March 5, 2025
Enforcement
- Recall Number
- Z-1208-2025
- Event ID
- 96270
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- INTELERAD MEDICAL SYSTEMS INCORPORATED
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 5, 2025
- Initiation Date
- February 5, 2025
- Classification Date
- February 21, 2025
- Address
- 800, Boul. de Maisonneuve Est, 12th floor, Montreal, N/A, N/A, Canada
Description
IntelePACS (Image Fusion Module) - InteleViewer
Reason
A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Code Info
UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
Distribution
Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Quantity
88