FDA Enforcement Class II Ongoing

IntelePACS (Image Fusion Module) - InteleViewer

Recall: Z-1208-2025 · Reported March 5, 2025

Enforcement

Recall Number
Z-1208-2025
Event ID
96270
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
INTELERAD MEDICAL SYSTEMS INCORPORATED
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 5, 2025
Initiation Date
February 5, 2025
Classification Date
February 21, 2025
Address
800, Boul. de Maisonneuve Est, 12th floor, Montreal, N/A, N/A, Canada

Description

IntelePACS (Image Fusion Module) - InteleViewer

Reason

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

Code Info

UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later

Distribution

Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia

Quantity

88