FDA Enforcement Class II Ongoing

Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000

Recall: Z-1208-2022 · Reported June 8, 2022

Enforcement

Recall Number
Z-1208-2022
Event ID
90027
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Bio-Rad Laboratories, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 8, 2022
Initiation Date
April 7, 2022
Classification Date
May 28, 2022
Address
4000 Alfred Nobel Dr, N/A, Hercules, CA, 94547-1803, United States

Description

Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000

Reason

The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CD¿ROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall.

Code Info

REF: 250-3020/Lot: AX90085, AY90085, AZ90085, BB90085, CB90085, DB90085; REF: 250-3000, UDI/DI: 00847817008276, Lot/Expiry: 64442741/ 2/28/2023, 64444208/ 2/28/2023, 64444218/ 3/26/2023, 64444715/ 3/31/2023, 64452929/ 3/31/2023, 64452922/ 5/31/2023

Distribution

US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR OUS: Canada, Spain, France, Italy, India, Vietnam, Philippines

Quantity

1893