FDA Enforcement Class II Ongoing

Colonoscope, Model Number CF-H180AL.

Recall: Z-1205-2024 · Reported March 6, 2024

Enforcement

Recall Number
Z-1205-2024
Event ID
93832
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 6, 2024
Initiation Date
January 11, 2024
Classification Date
February 29, 2024
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229, United States

Description

Colonoscope, Model Number CF-H180AL.

Reason

Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

Code Info

Model Number: CF-H180AL; UDI DI: 4953170339608; Serial Number: 2001426, 2001524, 2001614, 2002139, 2002212, 2002285, 2103817, 2104137, 2104581, 2105255, 2105976, 2107112, 2207880, 2207929, 2207940, 2309317, 2601078, 2601096, 2601255, 2601368, 2601700, 2702380, 2702460, 2702478, 2702805, 2703697, 2704366, 2704490, 2806996, 2807975, 2900004, 2900185, 2900249, 2908770, 2909110, 2909884;

Distribution

US Nationwide distribution.

Quantity

36