FDA Enforcement
Class II
Ongoing
Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation
Recall: Z-1205-2022
·
Reported June 8, 2022
Enforcement
- Recall Number
- Z-1205-2022
- Event ID
- 90213
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 8, 2022
- Initiation Date
- April 20, 2022
- Classification Date
- May 27, 2022
- Address
- 3 Chemin Du Cyclotron, N/A, Louvain-la-neuve, N/A, N/A, Belgium
Description
Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation
Reason
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed
Code Info
UDI-DI: (01)05404013801138 Serial Numbers: PAT108 (US), PAT112 (US), SAT116 (US). PAT107 (EU).
Distribution
Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.
Quantity
4 units