Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Recall: Z-1205-2022 · Reported June 8, 2022

Enforcement

Recall Number
Z-1205-2022
Event ID
90213
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 8, 2022
Initiation Date
April 20, 2022
Classification Date
May 27, 2022
Address
3 Chemin Du Cyclotron, N/A, Louvain-la-neuve, N/A, N/A, Belgium

Description

Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Reason

First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed

Code Info

UDI-DI: (01)05404013801138 Serial Numbers: PAT108 (US), PAT112 (US), SAT116 (US). PAT107 (EU).

Distribution

Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.

Quantity

4 units