FDA Enforcement Class II Terminated

CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.

Recall: Z-1204-2017 · Reported February 22, 2017

Enforcement

Recall Number
Z-1204-2017
Event ID
75942
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 22, 2017
Initiation Date
November 29, 2016
Classification Date
February 15, 2017
Termination Date
June 7, 2018
Address
511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States

Description

CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.

Reason

There is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type. This can occur when an order is received from the LIS without a sample type, requiring that the sample type be set in CentraLink based on the sample type of the test in the order.

Code Info

Product codes: 10485061, 10484212, 10485064, 10483439, 10471393, 10484552, 10814296, 10815474, 11220480, 10481936, 10481937, 10481938, 10482170, 10482936, 10483437, 10483438, 10483440, 10484211, 10484213, 10485060, 10485062, 10485063, 10703007, 10703008, 10711638, 10711639, 10711640, 10711641, 10711643, 10811596, 10811597, 10811598, 10811599, 10811601, 10811741, 10811742, 10811743, 10814298, 10814877, 10814880, 10814881, 10815694, 10816519, 10816520, 10816521, 10816522, 10816523, 10816524, 10817180, 10817181, 10819760, 10819943, 11219840, 11219841.

Distribution

Worldwide Distribution - USA (nationwide) to the states of : AL, AR, AZ, CA, CO, CT FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA and WV; and to the country of Canada and others.

Quantity

3,893 units