FDA Enforcement Class II Ongoing

Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001, STR-OR-001

Recall: Z-1202-2025 · Reported February 26, 2025

Enforcement

Recall Number
Z-1202-2025
Event ID
96091
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Dexcom, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
February 26, 2025
Initiation Date
January 27, 2025
Classification Date
February 20, 2025
Address
6340 Sequence Dr, N/A, San Diego, CA, 92121-4356, United States

Description

Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001, STR-OR-001

Reason

Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.

Code Info

All Serial Numbers/UDI: 00386270000583

Distribution

Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC, NH, FL, CA, NJ, PR, GA and the countries of Austria, Belgium, Canada, Switzerland, Germany, Spain, France, Ireland, Japan, Netherlands, Slovenia, United Kingdom, Australia, Czech Republic, Denmark, Finland, France, Hong Kong, Israel, Italy, Korea, Kuwait, New Zealand, Norway, Oman, Poland, Qatar, Saudi Arabia, Singapore, Slovakia, South Africa, Sweden, Turkey, United Arab Emirates.

Quantity

N/A