Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Recall: Z-1201-2014 · Reported March 19, 2014

Enforcement

Recall Number
Z-1201-2014
Event ID
67516
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mar Cor Purification
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
March 19, 2014
Initiation Date
June 27, 2012
Classification Date
March 13, 2014
Termination Date
August 22, 2014
Address
14550 28th Ave N, N/A, Minneapolis, MN, 55447-4817, United States

Description

Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Reason

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Code Info

serial numbers: 1289064, 1289069, 1289159, 1289259, 1289387, 1289650, 1289747, 1289826, 1289949, 1290088, 1290151, 1290333, 1290367, 1290480, 1290489, 1290845, 1290988, 1291042, 1291120, 1291233, 1291373, 1291378, 1291634, 1291777, 1291828, 1291933, 1292031, 1292114, 1292185, 1292185, 1292192, 1292391, 1292546, 1292597, 1292670, 1292674, 1292899, 1292960, 1293010, 1293039, 1293185, 1293190, 1293311, 1293379, 1293414, 1293687, 1293724, 1293771, 1293777, 1293953, 1293957, 1294093, 1294103, 1294171, 1294307, 1294362, 1294433, 1294516, 1294589, 1294727, 1294864, 1294999, 1295113, 1295137, 1295160, 1295186, 1295274, 1295391, 1295514, 1295622, 1295813, 1295871, 1296237, 1296241, 1296246, 1296484, 1296491, 1296582, 1296621, 1296650, 1296695, 1296806, 1297041, 1297046, 1297118, 1297231, 1297234, 1297372, 1297375, 1297394, 1297453, 1297484, 1297610, 1297655, 1297939, 1297943, 1298037, 1298210, 1298266, 1298270, 1298377, 1298407, 1298450, 1298457, 1298575, 1298619, 1298624, 1298704, 1298734, 1298797, 1298820, 1298879, 1298913, 1298952, 1298953, 1299037, 1299120, 1299144, 1299170, 1299241, 1299262, 1299285, 1299331, 1299393, 1299473, 1299512, 1299599, 1299655, 1299656, 1299732.

Distribution

nationwide. No distribution outside the US.

Quantity

130