FDA Enforcement Class II Ongoing

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia

Recall: Z-1199-2025 · Reported February 26, 2025

Enforcement

Recall Number
Z-1199-2025
Event ID
96155
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Augustine Temperature Management, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
February 26, 2025
Initiation Date
December 16, 2024
Classification Date
February 20, 2025
Address
15305 Minnetonka Blvd, N/A, Minnetonka, MN, 55345-1512, United States

Description

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia

Reason

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Code Info

UDI-DI UDI-DI 00855913001848, Serial Numbers: 1304384, 1304619, 1304620

Distribution

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

Quantity

3 units