FDA Enforcement
Class II
Ongoing
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
Recall: Z-1197-2025
·
Reported February 26, 2025
Enforcement
- Recall Number
- Z-1197-2025
- Event ID
- 96155
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Augustine Temperature Management, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- February 26, 2025
- Initiation Date
- December 16, 2024
- Classification Date
- February 20, 2025
- Address
- 15305 Minnetonka Blvd, N/A, Minnetonka, MN, 55345-1512, United States
Description
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
Reason
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Code Info
UDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297
Distribution
US Nationwide distribution in the states of TX, AL, NV; AU, NZ.
Quantity
4 units