FDA Enforcement
Class I
Terminated
A) Presource PBDS, Lap Chole, Kit, Clean Up; B) Presource PBDS, Greenwood LeFlore Hosp, Lap Chole, Kit
Recall: Z-1194-2013
·
Reported May 15, 2013
Enforcement
- Recall Number
- Z-1194-2013
- Event ID
- 64736
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cardinal Health, Medical Products & Services
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 15, 2013
- Initiation Date
- March 26, 2013
- Classification Date
- May 3, 2013
- Termination Date
- April 12, 2017
- Address
- 1430 Waukegan Road, Attn V. Mueller Qa, Park City, IL, 60085, United States
Description
A) Presource PBDS, Lap Chole, Kit, Clean Up; B) Presource PBDS, Greenwood LeFlore Hosp, Lap Chole, Kit
Reason
Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.
Code Info
A) Catalog Number: PG24LCGWG02; Lots: 153502, 162892, 170494, 180482, , 88701, 193548, 198163, 206986; B) Catalog Number: PG24LCGWG; Lots:153502, 162892, 170494, 180482, 188701, 193548, 198163, 206986
Distribution
Distributed in the states of AR, MS, SC, and TN.
Quantity
4,571 Kits total