FDA Enforcement Class II Ongoing

Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255

Recall: Z-1192-2023 · Reported March 8, 2023

Enforcement

Recall Number
Z-1192-2023
Event ID
91582
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
CareFusion 303, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 8, 2023
Initiation Date
January 24, 2023
Classification Date
March 1, 2023
Address
10020 Pacific Mesa Blvd, N/A, San Diego, CA, 92121-4386, United States

Description

Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255

Reason

Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed

Code Info

All PC Units, UDI-DI: 10885403801549, 10885403801532, with software versions 12.1.0, 12.1.2, 12.1.3, utilizing affected network cards and IP addresses

Distribution

US: FL, CA, OR, IL, WI, NJ, AZ, NV, CO, WY, NE, TX, MI, VA, OH, NC, LA, NM, AL, TN, AR, PA, CT, IN, KS, GA, SC, MD, UT, MA, DC, OK, IA, NY, WA, SD, ND, MN, KY, MO, WV, MS, DE, ME, MT, ID, NH, HI, AK, VT OUS: Canada, Qatar, Saudi Arabia,, Kuwait, United Arab Emirates, Israel, South Africa, Zimbabwe, Botswana, Australia, New Zealand

Quantity

83,185