FDA Enforcement Class II Ongoing

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Version: RayStation 12A, 12A Service Pack 1 and 2 and 2023B

Recall: Z-1191-2025 · Reported February 26, 2025

Enforcement

Recall Number
Z-1191-2025
Event ID
96156
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 26, 2025
Initiation Date
January 16, 2025
Classification Date
February 19, 2025
Address
Eugeniavagen 18c, N/A, Stockholm, N/A, N/A, Sweden

Description

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Version: RayStation 12A, 12A Service Pack 1 and 2 and 2023B

Reason

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Code Info

Lot Code: UDI: 0735000201054920220616, 0735000201067920221007, 0735000201073020230913 and 0735000201055620230630 GTIN: 07350002010549, 07350002010679, 07350002010730 and 07350002010556 Serial Numbers: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Revisions: RayStation 12A, 12A Service Pack 1 and 2 and 2023B Expiration Date: 2028-07-07

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.

Quantity

32