Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel¿ tube and outer housing.
Enforcement
- Recall Number
- Z-1190-2017
- Event ID
- 76236
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Greatbatch Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 22, 2017
- Initiation Date
- October 19, 2016
- Classification Date
- February 10, 2017
- Termination Date
- March 8, 2019
- Address
- 2300 Berkshire Ln N, N/A, Plymouth, MN, 55441-4575, United States
Description
Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel¿ tube and outer housing.
On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.
2481193, 2468380, 2599785, 2463309, 2468379, 2468398, 2468397, 2599516, 2671555, 2599517, 2670543, 2671558, 2734809, 2814828, 2814829, 2824496, 2858434, 2865404, 3030658, 2944397, 2977852, 2967214, 3000586, 3000590, 3030658, 3044689, 3071903, 3135658, 3088675, 3107504, 3188688, 3385844, 3188690, 3387046, 3397768.
Worldwide Distribution - US Nationwide in the states of IN and the countries of Canada, Austria, Germany, Japan.
1,441