FDA Enforcement Class II Terminated

Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel¿ tube and outer housing.

Recall: Z-1190-2017 · Reported February 22, 2017

Enforcement

Recall Number
Z-1190-2017
Event ID
76236
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Greatbatch Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 22, 2017
Initiation Date
October 19, 2016
Classification Date
February 10, 2017
Termination Date
March 8, 2019
Address
2300 Berkshire Ln N, N/A, Plymouth, MN, 55441-4575, United States

Description

Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel¿ tube and outer housing.

Reason

On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.

Code Info

2481193, 2468380, 2599785, 2463309, 2468379, 2468398, 2468397, 2599516, 2671555, 2599517, 2670543, 2671558, 2734809, 2814828, 2814829, 2824496, 2858434, 2865404, 3030658, 2944397, 2977852, 2967214, 3000586, 3000590, 3030658, 3044689, 3071903, 3135658, 3088675, 3107504, 3188688, 3385844, 3188690, 3387046, 3397768.

Distribution

Worldwide Distribution - US Nationwide in the states of IN and the countries of Canada, Austria, Germany, Japan.

Quantity

1,441