FDA Enforcement Class II Terminated

Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that the Calibrator set is not part of the IU but required for the assay.

Recall: Z-1187-2021 · Reported March 17, 2021

Enforcement

Recall Number
Z-1187-2021
Event ID
76372
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Tosoh Bioscience Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 17, 2021
Initiation Date
February 23, 2012
Classification Date
March 8, 2021
Termination Date
July 8, 2021
Address
3600 Gantz Rd, Grove City, OH, 43123-1895, United States

Description

Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that the Calibrator set is not part of the IU but required for the assay.

Reason

Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.

Code Info

Part Number 020391; Lot # B834868, Lot # BX34869, Lot # C234871, Lot # C434872 & Lot # C534873.

Distribution

US Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.

Quantity

34 boxess/2 sets/6 bottles