FDA Enforcement
Class II
Ongoing
Triton Canister Software Catalog Number: GAUS-2. Intended adjunct in the estimation of blood loss.
Recall: Z-1186-2023
·
Reported March 8, 2023
Enforcement
- Recall Number
- Z-1186-2023
- Event ID
- 91578
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 8, 2023
- Initiation Date
- January 25, 2023
- Classification Date
- February 27, 2023
- Address
- 1941 Stryker Way, N/A, Portage, MI, 49002-9711, United States
Description
Triton Canister Software Catalog Number: GAUS-2. Intended adjunct in the estimation of blood loss.
Reason
Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.
Code Info
UDI-DI: 00859506006029 Software Version: 2.8.4
Distribution
US Nationwide distribution.
Quantity
88 units