FDA Enforcement
Class II
Terminated
SL3 SOFT-TISSUE Laser (Model LR2002); Manuf: Discus Dental, LLC, 8550 Higuera Street, Culver City CA 90232. The SL3 is a soft-tissue diode laser intended to be used for dental procedures
Recall: Z-1186-2013
·
Reported May 8, 2013
Enforcement
- Recall Number
- Z-1186-2013
- Event ID
- 64621
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Discus Dental LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 8, 2013
- Initiation Date
- July 21, 2011
- Classification Date
- April 29, 2013
- Termination Date
- April 29, 2013
- Address
- 8550 Higuera St, Culver City, CA, 90232-2522, United States
Description
SL3 SOFT-TISSUE Laser (Model LR2002); Manuf: Discus Dental, LLC, 8550 Higuera Street, Culver City CA 90232. The SL3 is a soft-tissue diode laser intended to be used for dental procedures
Reason
The firm is initiating this recall of the SL3 Soft Tissue Laser equipped with a fiber optic Unifiber system which contains a fragile glass filament because the fiber broke near the hand piece.
Code Info
Model LR2002
Distribution
Nationwide Distribution including AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MO, MT NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, VA, and WA.
Quantity
150 units