FDA Enforcement Class II Ongoing

Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 3L Catalog Number: FG 12022. Intended adjunct in the estimation of blood loss.

Recall: Z-1185-2023 · Reported March 8, 2023

Enforcement

Recall Number
Z-1185-2023
Event ID
91578
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 8, 2023
Initiation Date
January 25, 2023
Classification Date
February 27, 2023
Address
1941 Stryker Way, N/A, Portage, MI, 49002-9711, United States

Description

Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 3L Catalog Number: FG 12022. Intended adjunct in the estimation of blood loss.

Reason

Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.

Code Info

UDI-DI: 10859506006101 Lot Numbers: M2112001, M2112003, M2202001, M2203001, M2204001, M2205001, M2206002

Distribution

US Nationwide distribution.

Quantity

541 units