Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher
Enforcement
- Recall Number
- Z-1185-2016
- Event ID
- 73403
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Medical Division of Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 23, 2016
- Initiation Date
- February 5, 2016
- Classification Date
- March 17, 2016
- Termination Date
- September 6, 2016
- Address
- 3800 E Centre Ave, Portage, MI, 49002-5826, United States
Description
Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher
The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we
The hydraulic jack assemblies (Part numbers: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as service parts on Stretchers assembled from 8/28/15 - 8/29/15
Distributed in DC and the states of AL, CA, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, and WV.
36