FDA Enforcement Class II Ongoing

Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.

Recall: Z-1181-2024 · Reported February 28, 2024

Enforcement

Recall Number
Z-1181-2024
Event ID
93988
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PTW-FREIBURG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 28, 2024
Initiation Date
February 8, 2024
Classification Date
February 22, 2024
Address
Lorracher Str. 7, N/A, Freiburg Im Breisgau, N/A, N/A, Germany

Description

Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.

Reason

Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.

Code Info

UDI-DI: EPTWS070031 Software VERIQA (S070031), version 2.0 and 2.1

Distribution

GA

Quantity

1 US; 65 worldwide