Quick Connects. Steris Corporation. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapters. For sterilant and rinse water used in endoscope washers.
Recall: Z-1180-2016
·
Reported March 23, 2016
Enforcement
- Recall Number
- Z-1180-2016
- Event ID
- 73009
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 23, 2016
- Initiation Date
- December 4, 2015
- Classification Date
- March 16, 2016
- Termination Date
- August 26, 2016
- Address
- 5960 Heisley Rd, Mentor, OH, 44060-1834, United States
Description
Quick Connects. Steris Corporation. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapters. For sterilant and rinse water used in endoscope washers.
Reason
The firm discovered that one of the hysteroscope models listed for processing in four existing Quick Connects has one inlet port rather than two inlet ports as indicated in the Quick Connect processing instructions.
Code Info
All Lots. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT.
Distribution
Worldwide distribution. US nationwide, Australia, Bahrain, Bulgaria, Canada, Greece, Italy, Malaysia, Mexico, Qatar, Russia, Spain, and United Kingdom.
Quantity
809 units