FDA Enforcement Class II Ongoing

GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes

Recall: Z-1179-2024 · Reported February 28, 2024

Enforcement

Recall Number
Z-1179-2024
Event ID
93882
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
W.L. Gore & Associates, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 28, 2024
Initiation Date
January 19, 2024
Classification Date
February 22, 2024
Address
301 Airport Rd, N/A, Elkton, MD, 21921-4130, United States

Description

GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes

Reason

Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary procedure, cholangitis, cholecystitis, and/or pain. The GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis is a flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. The endoprosthesis is radially compressed and secured onto the distal end of a catheter-based delivery system for deployment. The catheter provides a means for accessing and implanting the GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis at the target site in the biliary tract.

Code Info

(1) UDI-DI 00733132638802, Serial Numbers 27063943, 27063944, 27063949, 27063950, 27063952, 27063953; (2) UDI-DI 00733132638871, Serial Numbers 27063379, 27063380, 27063381, 27063382, 27063383, 27063384, 27063385, 27063386, 27063387, 27063426, 27063427, 27063428

Distribution

US Nationwide distribution in the state of Georgia.

Quantity

24