FDA Enforcement Class II Terminated

VentStar Basic (N) 180- Drager Breathing Circuits/Anesthesia Sets Order Number: MP00353 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.

Recall: Z-1178-2021 · Reported March 17, 2021

Enforcement

Recall Number
Z-1178-2021
Event ID
87267
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Draeger Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 17, 2021
Initiation Date
January 29, 2021
Classification Date
March 5, 2021
Termination Date
November 1, 2021
Address
3135 Quarry Rd, Telford, PA, 18969-1042, United States

Description

VentStar Basic (N) 180- Drager Breathing Circuits/Anesthesia Sets Order Number: MP00353 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.

Reason

Y-piece may become detached from the ventilation hose either before or during the ventilation process, a leak will be caused that can hinder ventilation of the patient.

Code Info

All lots

Distribution

US Nationwide distribution.

Quantity

1,525 units