FDA Enforcement Class II Ongoing

Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.

Recall: Z-1172-2024 · Reported February 28, 2024

Enforcement

Recall Number
Z-1172-2024
Event ID
93855
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
FUJIFILM Healthcare Americas Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 28, 2024
Initiation Date
January 3, 2024
Classification Date
February 21, 2024
Address
81 Hartwell Ave Ste 300, N/A, Lexington, MA, 02421-3160, United States

Description

Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.

Reason

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Code Info

UDI-DI 00854904006015 Software versions 6.2.1, 6.2.0, 6.1.1, 6.1.0, 6.0.1, and 6.0.0

Distribution

US domestic distribution nationwide. International distribution to Israel.

Quantity

8