Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.
Enforcement
- Recall Number
- Z-1171-2016
- Event ID
- 73129
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Sekisui Diagnostics P.E.I. Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 23, 2016
- Initiation Date
- January 22, 2016
- Classification Date
- March 14, 2016
- Termination Date
- August 9, 2016
- Address
- 70 Watts Ave, Charlottetown, N/A, Canada
Description
Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.
Crystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover
Catalog Number: 506-10 Lot Codes 47677 47694 47649 47650 47662 47663 48253 48385 48386 Catalog Number: 506-30 Lot Codes: 47695 48294
Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.
14,407 kits.