FDA Enforcement
Class II
Terminated
Product 69 consists of all product under product code: JDI and same usage: Item no: 785301701 VERSYS CEMENTED FEM STEM 785301521 VERSYS CEMENTED FEM STEM 15X140MM X-OFF 785301601 VERSYS CEMENTED FEM STEM 16X145MM For use in total or hemi hip arthroplasty
Recall: Z-1167-2016
·
Reported March 23, 2016
Enforcement
- Recall Number
- Z-1167-2016
- Event ID
- 72900
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Manufacturing B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 23, 2016
- Initiation Date
- January 11, 2016
- Classification Date
- March 12, 2016
- Termination Date
- April 13, 2017
- Address
- Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715, United States
Description
Product 69 consists of all product under product code: JDI and same usage: Item no: 785301701 VERSYS CEMENTED FEM STEM 785301521 VERSYS CEMENTED FEM STEM 15X140MM X-OFF 785301601 VERSYS CEMENTED FEM STEM 16X145MM For use in total or hemi hip arthroplasty
Reason
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Code Info
lot no.: 370967 00370967N 62279438 62012576 62012576N
Distribution
US Nationwide Distribution
Quantity
13